I Just Got My Probe Recalibrated, Now What?

I just got my probe recalibrated, so now I just hook it up to the readout, load the coefficients and start measuring. I don’t have to worry about the probe’s calibration until it’s recalibrated.

Surprisingly, this is exactly how many people handle their probes’ calibrations. Following this practice saves a little time, but the cost can be severe. Recently, a company lost over a million dollars of product and six months of calibration work. The culprit was simply a PRT probe that shifted, probably during shipping, and wasn’t checked before being returned to service.

I hope this article will give you some ideas on how to make your temperature measurement program bulletproof by closing the loop on your probe calibrations.

What does “closing the loop” mean?

“Closing the loop” is completing the series of steps a probe should pass through as it is calibrated and used. In other words, there is more involved in managing a probe’s calibration than just sending the probe out for calibration every year.

The first step is a measurement taken to capture the status of the probe before it is boxed up and sent into shipping. This measurement closes the loop on the work the probe has done. When the probe is received back from calibration, another measurement is taken to verify that the probe didn’t shift during return shipment. Both the send-off measurement and return measurement are compared with the calibration report. This enables you to monitor the status of the probe, and it also provides a crosscheck between the calibration lab and your lab.

Measuring a probe before and after shipment may be the most important measurements in the overall calibration loop. Many probes and instruments are damaged beyond repair while in shipping. Make sure your work is closed out before shipping a probe.

How do I know if the probe is in tolerance?

Just because the probe was calibrated and given a calibration certificate doesn’t mean it was in tolerance. This might sound obvious, but I have had this conversation with enough customers to know that it is a common misconception. Someone who is surprised by this always follows up with the question “How do I know if the probe is in tolerance?”

Basically, the tolerance status of a probe is calculated by comparing the new calibration with the previous calibration. We answered this question in more detail in a previous Random News. Please refer to Random News Number 8.

How can “out of tolerance” be avoided?

The goal of all measurement programs is to make sure that everything always works and nothing is out of tolerance. Of course, the better the measurement program, the less often instruments are out of tolerance. Measuring a probe as it is sent out for calibration and when it is received from calibration is good, but the probe also needs to be monitored as it is being used.

One of the difficulties in measuring temperature is that probes are typically very fragile so their performance is very dependent on care and handling. Manufacturers may spec a probe’s long-term drift, but the only sure way to know the performance of a probe is by implementing interim checks and control charts (see the sample control chart in Figure 1). Whenever interim checks are mentioned, someone always says “Who has the time?” My answer is “Who has the time to recall six months or a year’s worth of work?” In the situation mentioned earlier, a 30-minute check every week would have saved the company over a million dollars and six months of work.

How often should the probe be checked?

There are no established rules that define how often an interim check should be done. One approach to establishing an interval is to decide the largest recall interval you are comfortable with and set the interim check interval accordingly. For example, if you calibrate 10 instruments a week with your temperature probe, then a weekly interim check may be sufficient. The largest risk would be the recall of 10 instruments. A daily interim check would limit the risk to recalibration of two instruments.

At each interim check, record the data on a spreadsheet or in a lab book. Graph the data. This is the beginning of a control chart. Add control lines to the chart to indicate how much the probe can drift before it exceeds the limits set in the calibration program (see Figure at right). The sample control chart shows the data of a probe measured about every seven days. On 4/28/06 the probe’s shift was verified to be about 0.006 °C, which exceeds the process limit. Only four days of cal work may have to be recalled due to the shift.

What temperature points should be measured?

This is an essential question to ask because, depending on the situation, many different temperatures can be used. Mainly, the temperature point(s) should be fairly simple to achieve and, importantly, the temperature point(s) must be repeatable. Let’s go over a few options.

The triple-point of water (TPW), 0.010 °C, is most often used for closing the loop in probe calibrations. At Fluke Calibration, the TPW is measured multiple times in every platinum resistance thermometer (PRT) calibration to monitor the probe as it is calibrated.

The TPW is very important in PRT calibrations because it can be used to troubleshoot several types of problems that affect PRT performance. Not only is the TPW included in every ITS-90 calibration, it is used in every ITS-90 measurement as well. The ITS-90 was designed to allow the user to update the probe’s RTPW (resistance at the triple-point of water). Updating the RTPW coefficient can also help reduce errors due to probe drift. For these reasons, the TPW is the first choice for a quick check to verify and close the loop on a PRT.

 

  • ホーム
  • 製品
    • 新製品情報
    • 電気校正
      • 電気標準器
      • 電気校正器
      • ベンチ・マルチメーター
      • 電気校正ソフトウェア
    • RF 校正
      • RF リファレンス・ソース
      • RF 校正アクセサリー
      • RF 校正ソフトウェア
    • データ収集とテスト装置
      • ベンチ・マルチメーター
      • データ収集
      • ファンクションおよび波形ジェネレーター
      • データ収集ソフトウェア
    • 温度校正
      • ITS-90 温度標準器
        • ITS-90 定点セル
        • 標準白金抵抗温度計(SPRT)
        • メンテナンス装置
        • 液体窒素式比較校正装置
        • 抵抗ブリッジ
        • 標準抵抗器
      • 校正バス
        • 小型校正バス
        • 標準バス
        • 特殊アプリケーション用バス
        • バス用アクセサリー
        • バス・コントローラー
        • バス用温度媒体
        • カスタム温度校正バス
      • 工業用温度校正器
        • フィールド・メトロロジー・ウエル
        • メトロロジー・ウェル
        • ハンドヘルド・ドライウェル校正器
        • フィールド・ドライウエル
        • 超小型バス
        • 赤外線校正器
        • 熱電対炉
        • デュアル・ブロック・ドライウエル
        • ゼロ点ドライウエル
      • プローブ/センサー
        • 白金抵抗温度計 (PRT)
        • サーミスター
        • 標準熱電対
      • デジタル温度指示計
      • マルチファンクション校正器
      • 温度校正ソフトウェア
    • 湿度校正
      • 湿度校正器
      • 湿度データ・ロガー/モニター
    • 圧力校正
      • ピストン・ゲージ
        • NMI ピストン・ゲージ
        • 絶対圧ピストン・ゲージ
        • 高圧力ガス・ピストン・ゲージ
        • 油圧式ピストン・ゲージ
        • ピストン・ゲージ用アクセサリー
      • 圧力コントローラー/校正器
        • 低圧力コントローラー/校正器
        • 空気式圧力コントローラー/校正器
        • 高圧空気式コントローラー/校正器
        • 油圧式コントローラー/校正器
      • 圧力モニター
        • 低圧力計
        • デジタル圧力計
        • 基準圧力モニター
      • 重錘形圧力計
        • 空気式デッドウェイト・テスター
        • 油圧式デッドウェイト・テスター
        • 液圧式デッドウェイト・テスター
        • 高圧油圧式デッドウェイト・テスター
        • デッドウェイト・テスター用アクセサリー
      • 手動校正器、モニター
        • 圧力校正器
        • 空気圧の制御
        • 液体圧比較器/ポンプ
      • Handheld Pressure Calibrators
      • エア・データ・テスト
      • ラボ環境モニター
      • 圧力校正アクセサリ
      • 圧力校正カスタム・システム
      • 圧力校正ソフトウェア
    • 流量校正
      • 気体流量標準器
      • 気体流量アクセサリ
      • GFS 質量流量一次標準
      • 流量校正ソフトウェア
    • プロセス校正ツール
      • 温度校正器
        • ハンドヘルド温度校正器
        • ドライブロック校正器およびマイクロバス
        • 高精度デジタル温度計
        • 温度プローブ
        • 赤外線温度計校正器
        • データ・ロギング機能付き湿温度計
      • 圧力校正器
        • デジタル圧力校正器
        • ハンドヘルド圧力校正器
        • 重錘形圧力計
        • 高精度デジタル圧力計
        • 校正用ハンド・ポンプ
      • マルチファンクション校正器
      • mA ループ校正器
      • プロセス校正ソフトウェア
    • 校正ソフトウェア
      • MET/CAL ソフトウェア
      • MET/CAL Support
      • 資産管理ソフトウェア
      • 温度校正ソフトウェア
      • 圧力校正ソフトウェア
      • 流量校正ソフトウェア
      • 機械/寸法校正ソフトウェア
    • サービスとサポート
    • すべての校正機器
  • ご購入について
    • 購入窓口
    • 見積依頼
    • デモ依頼
    • 相談/問い合わせ
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • ニュース
    • プレスリリース
    • プロモーション
    • 業界リンク
    • Metrology Salary Survey
  • トレーニングとイベント
    • 展示会/セミナー
    • トレーニング・コース
    • User Group Meetings
    • ウェブセミナー
      • ライブ・セミナー
      • 過去のセミナー(アーカイブ)
  • カタログ・資料
    • Education Hub
    • About Calibration
    • 参考資料
    • 製品資料
    • 製品マニュアル(ユーザー・ガイド)
    • Videos and Virtual Demos
    • Blog
  • サービスとサポート
    • Service Request (RMA)
    • Service Plans
    • ナレッジベース
    • 認定校正
    • サービスセンター
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • 製品マニュアル(ユーザー・ガイド)
    • 製品の登録
    • 安全データ・シート(SDS)
    • 安全性および製品に関するお知らせ
    • ソフトウェア・ダウンロード
  • フルーク・キャリブレーション
    • お問い合わせ先
    • 所在地
    • キャリア
    • フルーク・キャリブレーションとは
      • Hart Scientific:温度校正
      • DH Instruments:圧力/流量校正
      • Pressurements:圧力校正
      • Ruska:圧力校正
    • 関連企業
    • 行動規範とコンプライアンス
    • よくある質問
ホーム|カタログ・関連資料|参考資料|温度校正|研究論文/記事|I Just Got My Probe Recalibrated, Now What?

サイトの改善のため、フィードバックをお寄せください。
フィードバック・フォーム≫
©1995-2021 Fluke Corporation
             

Secondary menu

  • 免責事項
  • プライバシー・ステートメント
  • ウェブマスター
  • Terms of Use
  • Terms and Conditions of Sale