How a pressure balance works

Imagine for a moment you are Bob. Bob lives in New York and recently purchased a deadweight tester from a gentleman in South Carolina. The deadweight tester came with a certificate of calibration, but the instrument appears to read 0.5 to 0.6 psi off when calibrating his 0.05% 1,000 psi gauges. Bob scratches his head, then double checks the gauges with the older equipment he is replacing. The gauges read well within tolerance on his old equipment. Bob is not happy. Can you blame him?
 
Bob spent a lot of money on his new deadweight tester and it is effectively useless. Or is it? Bob tells his friend Jeff about his lousy luck. Jeff asks Bob if he is correcting for local gravity. “Yeah right,” Bob laughs. But Jeff isn’t laughing. “You know differences in local gravity can easily account for the error you’re seeing, right?”
 
Bob overlooked the basics of pressure calibration. He assumed that a deadweight tester that is certified in South Carolina would still be in tolerance in New York, and it came back to bite him. Don’t worry, though. There is still hope for Bob and his deadweight tester, but I’m getting ahead of myself.
 
A deadweight tester is basically a pressure balance. A pressure balance is a very elegant solution to the pressure measurement problem. So how does it work?
 

Pressure is fundamental

 
Actually, it’s not. Pressure cannot be described by a single variable like length, time, temperature or amount of substance. Pressure is force per unit area. So what we have to do to get pressure is combine the units of length, time, and mass. The unit of pressure is the Pascal (Pa), which is defined as 1 kg*m/second2. That’s actually a very small amount of pressure. Atmospheric pressure is about 100,000 times larger than 1 Pascal. But sometimes little things can mean a lot.

How a pressure balance works

 
When we use a pressure balance, we are looking for the equilibrium between a downward force and an upward force. The downward force is that of a mass being accelerated by gravity. The upward force is directed by the pressure we are measuring onto the effective area of the piston, which is pressed downward by the mass. When these two forces are equal, they counter balance each other.
 
When the forces counter balance each other in this way, the mass will float. The trick is to know the exact force applied by the mass and the effective area in contact with the pressure we are trying to measure. Then we can calculate the pressure.

The floating piston

 
The beauty of this method is related to the floating mechanism. When the piston floats, it compensates automatically for a number of potential errors in our measurements. For example, we could have a leak, but instead of causing the pressure to drop the piston floats a little lower. The temperature of the medium could rise, but instead of causing the pressure to rise the piston floats a little higher. However, if the pressure actually rises or lowers then the piston no longer floats.
 
You may be thinking “Wait a minute, what about friction? Won’t friction between the piston and the inside walls of the chamber cause a measurement error?” You’re correct, friction is important. But the solution for friction is as elegant as the pressure balance itself. We have to spin the piston. When the piston spins, it ensures that friction will be minimized.

Deadweight tester assumptions

The type of pressure balance commonly used in industry is called a deadweight tester. Deadweight testers define pressure based on a couple of assumptions. The two big ones are:

1. Piston cylinder temperature
2. Local gravity
 
The pressure value determined by a deadweight tester is the result of summing the pressure values assigned to the loaded weights. Typically the weights will have a value stamped or engraved on them. The weights are called platters. When you take that platter, put it on the matching piston, and you float it, you are going to define nominal pressure as the value stamped on that platter.
 
However, to maintain the certified accuracy of the deadweight tester, you have to keep the environmental temperature within the stated limits of its assigned reported temperature, and you have to be at the same local gravity that the platters were manufactured and trimmed for. It is local gravity that caused big problems for Bob, and it is the single most significant source of uncertainty for a deadweight tester.

The gravity of the situation

Many people don’t realize that standard international gravity is defined at 45 degrees latitude north or south of the equator at sea level. It has a value of 9.80665 m/s2. Phoenix Arizona is closer to the equator than 45 degrees latitude, and the value of local gravity is 9.79474 m/s2. The difference is more than 1200 ppm (0.12%)! An industrial deadweight tester is typically specified to around 150 ppm, and with quality hardware and the right corrections you can even do much better than that. But if you fail to take into account local gravity, you don’t stand a chance.

 
So, what about Bob? It turns out Bob could apply a correction to ensure that his values are accurate. Using a ratio of his local gravity to the gravity his deadweight tester was originally designed for, he can make corrections to the pressure stamped on his platters. Of course there is more to life than pressure measurement, but if something is worth doing, it probably worth doing right.
Ron Ainsworth, Temperature Business Unit Manager at Fluke Corporation
Posted by Ron Ainsworth on 6月 27, 2012
Ron Ainsworth is the Business Manager for Process Calibration Tools at Fluke Calibration. After graduating with a degree in physics in 1998, he started in the primary temperature calibration laboratory in American Fork Utah. He has since had roles as a laboratory manager and marketing manager at... Full Bio
  • ログイン(登録)してコメントを投稿
  • Printer-friendly version »
  • ホーム
  • 製品
    • 新製品情報
    • 電気校正
      • 電気標準器
      • 電気校正器
      • ベンチ・マルチメーター
      • 電気校正ソフトウェア
    • RF 校正
      • RF リファレンス・ソース
      • RF 校正アクセサリー
      • RF 校正ソフトウェア
    • データ収集とテスト装置
      • ベンチ・マルチメーター
      • データ収集
      • データ収集ソフトウェア
    • 温度校正
      • ITS-90 温度標準器
        • ITS-90 定点セル
        • 標準白金抵抗温度計(SPRT)
        • メンテナンス装置
        • 液体窒素式比較校正装置
        • 抵抗ブリッジ
        • 標準抵抗器
      • 校正バス
        • 小型校正バス
        • 標準バス
        • 特殊アプリケーション用バス
        • バス用アクセサリー
        • バス・コントローラー
        • バス用温度媒体
      • 工業用温度校正器
        • フィールド・メトロロジー・ウエル
        • メトロロジー・ウェル
        • ハンドヘルド・ドライウェル校正器
        • フィールド・ドライウエル
        • 超小型バス
        • 赤外線校正器
        • 熱電対炉
        • デュアル・ブロック・ドライウエル
        • ゼロ点ドライウエル
      • プローブ/センサー
        • 白金抵抗温度計 (PRT)
        • サーミスター
        • 標準熱電対
      • デジタル温度指示計
      • マルチファンクション校正器
      • 温度校正ソフトウェア
    • 湿度校正
      • 湿度校正器
      • 湿度データ・ロガー/モニター
    • 圧力校正
      • ピストン・ゲージ
        • NMI ピストン・ゲージ
        • 絶対圧ピストン・ゲージ
        • 高圧力ガス・ピストン・ゲージ
        • 油圧式ピストン・ゲージ
        • ピストン・ゲージ用アクセサリー
      • 圧力コントローラー/校正器
        • 低圧力コントローラー/校正器
        • 空気式圧力コントローラー/校正器
        • 高圧空気式コントローラー/校正器
        • 油圧式コントローラー/校正器
      • 圧力モニター
        • 低圧力計
        • デジタル圧力計
        • 基準圧力モニター
      • 重錘形圧力計
        • 空気式デッドウェイト・テスター
        • 油圧式デッドウェイト・テスター
        • 液圧式デッドウェイト・テスター
        • 高圧油圧式デッドウェイト・テスター
        • デッドウェイト・テスター用アクセサリー
      • 手動校正器、モニター
        • 圧力校正器
        • 空気圧の制御
        • 液体圧比較器/ポンプ
      • Handheld Pressure Calibrators
      • エア・データ・テスト
      • ラボ環境モニター
      • 圧力校正アクセサリ
      • 圧力校正カスタム・システム
      • 圧力校正ソフトウェア
    • 流量校正
      • 気体流量標準器
      • 気体流量アクセサリ
      • GFS 質量流量一次標準
      • 流量校正ソフトウェア
    • プロセス校正ツール
      • 温度校正器
        • ハンドヘルド温度校正器
        • ドライブロック校正器およびマイクロバス
        • 高精度デジタル温度計
        • 温度プローブ
        • 赤外線温度計校正器
        • データ・ロギング機能付き湿温度計
      • 圧力校正器
        • デジタル圧力校正器
        • ハンドヘルド圧力校正器
        • 重錘形圧力計
        • 高精度デジタル圧力計
        • 校正用ハンド・ポンプ
      • マルチファンクション校正器
      • mA ループ校正器
      • プロセス校正ソフトウェア
    • 校正ソフトウェア
      • MET/CAL ソフトウェア
      • MET/CAL Support
      • 資産管理ソフトウェア
      • 温度校正ソフトウェア
      • 圧力校正ソフトウェア
      • 流量校正ソフトウェア
      • 機械/寸法校正ソフトウェア
    • サービスとサポート
    • すべての校正機器
  • ご購入について
    • 購入窓口
    • 見積依頼
    • デモ依頼
    • 相談/問い合わせ
    • Certified Pre-Owned Equipment
    • General Services Administration (GSA)
    • Financing Program
    • National Stock Numbers (NSNs)
    • Payment Options and Tax Info
  • ニュース
    • プレスリリース
    • プロモーション
    • 業界リンク
    • Metrology Salary Survey
  • トレーニングとイベント
    • 展示会/セミナー
    • トレーニング・コース
    • User Group Meetings
    • ウェブセミナー
      • ライブ・セミナー
      • 過去のセミナー(アーカイブ)
  • カタログ・資料
    • Education Hub
    • About Calibration
    • 参考資料
    • 製品資料
    • 製品マニュアル(ユーザー・ガイド)
    • Videos and Virtual Demos
    • Blog
  • サービスとサポート
    • Service Request (RMA)
    • Service Plans
    • ナレッジベース
    • 認定校正
    • サービスセンター
    • Calibration Certificates
    • Community Forum
    • My MET/SUPPORT
      • Activate
      • Procedures
      • Software
      • Technical Bulletins
      • Priority Support
      • Manuals
    • 製品マニュアル(ユーザー・ガイド)
    • 安全データ・シート(SDS)
    • 安全性および製品に関するお知らせ
    • ソフトウェア・ダウンロード
  • フルーク・キャリブレーション
    • お問い合わせ先
    • 所在地
    • キャリア
    • フルーク・キャリブレーションとは
      • Hart Scientific:温度校正
      • DH Instruments:圧力/流量校正
      • Pressurements:圧力校正
      • Ruska:圧力校正
    • 関連企業
    • 行動規範とコンプライアンス
    • よくある質問
ホーム|Blog|How a pressure balance works
©1995-2022 Fluke Corporation
             

Secondary menu

  • 免責事項
  • プライバシー・ステートメント
  • Terms of Use
  • Terms and Conditions of Sale